The Impact of Skin Pigmentation on Blood Oxygen Measurement Accuracy and Analysis of Calibration Methods
Abstract
This study uses COMSOL Multiphysics to create an optical model incorporating layered skin structures to investigate the impact of melanin concentration on light absorption characteristics at red and infrared wavelengths. The dynamic calibration algorithm is proposed to enhance measurement accuracy. Simulation results show that under high melanin concentrations, red light transmission intensity decreases by 23.5%, leading to an R-value deviation of 6.8%. By integrating the Individual Typology Angle for skin tone quantification and applying multiple linear regression, the root mean square error (RMSE) is reduced from 5.44% to 1.2%. This research provides theoretical support for optimizing optical design and skin-tone calibration in pulse oximeters, contributing to improved health monitoring equity for darker-skinned populations.
References
- Ren, Y Z. (2024), CS32A010 oximeter based solution[J]. China Integrated Circuit, 300: 79-83.
- Sohn R. (2022), The Fatal Flaw of Pulse Oximetry[J]. Biomedical.
- Guo C Y, et al. (2023), Calibrating Oxygen Saturation Measurements for Different Skin Colors Using ITA[J]. IEEE Sensors Magazine, 23(15): 16994-17001.
- Ma, J. (2023), Correction of oxygen saturation R-value curves based on the Lambert-Beer law[J]. Measurement Science and Technology, 67(1): 23-28.
- Chen, N. (2023), Tracking Blood Oxygen - Spectroscopy at Your Fingertips[J]. University Chemistry, 38(7): 136-141.
- Chen, H. (2023), Oximeter: CN116919398A[P].
- Shao, S H. (2024), A study of calibration methods for pulse oximetry[J]. Metrology and test technology, 50(12): 128-130.
- Y 0784-2010, Basic Safety Requirements for Medical Pulse Oximetry Equipment[S].
- Chen, Y. Li, K Y. (2021), A study of three-wavelength oximetry calibration curves[J]. Beijing Biomedical Engineering, 40(5): 499-504.
- Chen, N,You, T. (2021), Analysis of factors affecting the accuracy of pulse oximetry measurement [J]. China Medical Engineering,2021,29(5):13-16.
- D. R. Seshadri et al, (2020), "Wearable Sensors for COVID-19: A Call to Action to Harness Our Digital Infrastructure for Remote Patient Monitoring and Virtual Assessments, " Frontiers in Digital Health, vol. 2, pp. 1-11, Jun.