Enhancing Kalman Filter Performance for PPG Signal Denoising: Adaptive Integration of Accelerometer and Gyroscope Data

Weixian Ao*
Canada British Columbia International Schools, Chengdu, Sichuan Province, China, 610000

* Corresponding author: [email protected]

Abstract

The monitoring of oxygen saturation (SpO2) for premature babies carries great importance, due to the highly required accuracy of oxygen therapy, which is used to prevent premature babies’ dyspnea. Motion artifacts, however, can influence the quality of the photo-plethysmograph (PPG) signal, causing imprecision. Through this study, the accuracy of PPG detection for newborns can be improved by applying the Kalman filter to restitute motion artifacts by using an accelerator and a gyroscope to build reference signals and adaptively adjust the parameters of the Kalman filter. A quantitative study is used in this investigation with relative simulations of Heartbeat Rate (HR) estimation based on filtered PPG signals. The results, on average, show an absolute error of 1.69 BPM, a relative error of 1.82%, and a standard deviation of 9.37 BPM. The great accuracy of the algorithm can improve SpO2 monitoring for newborns, further lowering the working pressure for medical workers. On the other hand, the standard deviation is relatively high, indicating that further work, such as improving the preprocess of signals or introducing machine learning, should be paid attention to.

Keywords: Pulse Oximeter; PPG Signal; Motion Artifact; Accelerator; Gyroscope; Kalman Filter
Published: January 24, 2025
DOI: 10.54254/2755-2721/2025.20638
Volume: ACE Vol.133
pp. 142-148
Download PDF

References

  1. Majumder, S., Mondal, T., Deen, M.J. (2017) Wearable sensors for remote health monitoring. J. Sensors, 130: 1-45.
  2. Brostowicz, H.M., Rais-Bahrami, K. (2010) Oxygen saturation monitoring in the Neonatal Intensive Care Unit (NICU): Evaluation of a new alarm management. J. Neonatal-Perinatal Med, Vol. 3: 201-205.
  3. Subhagya D.S., Keshava. M.C., Aruna N., Janardhan, L., Ramakrishna H.S. (2017) Case study on measurement of SpO2 from PPG signals in the presence of motion artifacts. In: International Conference on Recent Advances in Electronics and Communication Technology. pp. 318-323.
  4. Hertzman, A.B. (1938) The blood supply of various skin areas as estimatedby the photoelectric plethysmograph. J. Physiol.-Legacy Content,vol. 124, no. 2: 328–340.
  5. B.-L. Lee, B.-L. Lee, W.-Y. Chung, (2015) Smartwatch-based driver alert-ness monitoring with wearable motion and physiological sensor, in Proc.37th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc., vol. 1., pp. 6126–6129.
  6. Poets, C.F., Stebbens, V.A. (1997) Detection of movement artifact in recorded pulse oximeter saturation. J. EUROPEAN JOURNAL OF PEDIATRICS.,156 (10): 808-811.
  7. Cho, J.H., Kim, J.C., Yoon, G.W. (2014) Robust design of pulse oximeter using dynamic control and motion artifact detection algorithms. J. JOURNAL OF ELECTRICAL ENGINEERING & TECHNOLOGY., Vol. 5: 1780-1787.
  8. Zahari, M., Lee, D.S., Darlow, B,A. (2016) Algorithms that eliminate the effects of calibration artefact and trial-imposed offsets of Masimo oximeter in BOOST-NZ trial. J. Clin Monit Comput., 30(5): 669-78.
  9. Banik, P.P., Hossain, S., Kwon, T., Kim, H., Kim, K. (2020) Development of a wearable reflection-type pulse oximeter system to aquire clean PPG signals and measure pulse rate and SpO2 with and without finger motion. J. Electronics., Vol. 9.
  10. Harrison, W., Lim, J., Singer, E. (1986) A new application of adaptive noise cancellation. J. IEEE TRANSACTIONS ON ACOUSTICS SPEECH AND SIGNAL PROCESSING., Vol 34: 21-27.
  11. Wu, C.H., Lee, J.H., Yang, C.C.H. (2020) Improving the diagnostic ability of the sleep apnea screening system based on oximetry by using physical activity data. J. JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING., Vol. 40: 858-867.
  12. Yang, D., Cheng, Y.Q., Peng, Y.H. (2018) A novel adaptive spectrum noise cancellation approach for enhancing heartbeat rate monitoring in a wearable device. J. IEEE ACCESS., Vol. 6: 8364-8375.
  13. Lee, B., Han, J., Yi, W.J. (2010) Improved elimination of motion artifacts from a photoplethysmographic signal using a Kalman smoother with simultaneous accelerometry. J. PHYSIOLOGICAL MEASUREMENT., Vol. 32: 1585-1603.