Research Article

Design of Multi Sensor Integrated Oximeter Based on STM32

Zhixuan Wang*
School of Automation, Wuhan University of Technology, Wuhan, Hubei Province, China, 430070

* Corresponding author: [email protected]

Abstract

With the growing emphasis on physical health monitoring, blood oxygen data has become a key indicator of overall health. Therefore, the design of oximeter is crucial for meeting people's daily health needs. Traditional single-function oximeter cannot meet people's requirements for obtaining multiple health data at the same time. Therefore, this article designs an oximeter based on STM32 microcontroller. This oximeter integrates additional functions such as heart rate detection, temperature detection, and step detection on the basis of blood oxygen measurement, and can provide more data than a simple oximeter. The system utilizes STM32 as the main control chip, MAX30102 as the heart rate and blood oxygen sensor, DS18B20 as the temperature sensor, and ADXL345 as the step detection sensor. Experimental results show that the multi-sensor oximeter provides accurate temperature, step count, and blood oxygen data, with the latter closely matching the measurements obtained from an Apple Watch, thus meeting the design requirements.

Keywords: adaptive; oximeter; circuit design; STM32
Published: January 24, 2025
DOI: 10.54254/2755-2721/2025.20606
Volume: ACE Vol.133
pp. 96-104
Download PDF

References

  1. Luo, Z., Guo, Q., Zheng, M., Yu, H., Zhu, C., Zhang, L., & Zhang, W. (2019). Design and implementation of an ear-mounted pulse oximeter. China Medical Device Information, 25(03), 15-18+41. https://doi.org/10.15971/j.cnki.cmdi.2019.03.005 (in Chinese)
  2. Liu, M., & Li, X. (2021). Design of a clip-on pulse oximeter based on a photodiode sensor. Electronic Testing, 47(15), 21-22+50. https://doi.org/10.16520/j.cnki.1000-8519.2021.15.006 (in Chinese
  3. Pang, G., & Ma, C. (2014). A neo-reflective wrist pulse oximeter. IEEE access, 2, 1562-1567.
  4. Shakhreet, B. Z., & Shakhreet, M. Z. (2018). Design of pulse oximetry system. J. Basic Appl. Res. Int, 18, 3457.
  5. Bai, P., Liu, Q., Duan, F., An, Q., de Rooij, N. F., Li, N., & Zhou, G. (2017). Design of a wearable pulse oximetry system based on MAX30102. Laser & Infrared, 47(10), 1276-1280.
  6. He, Z., Wang, J., Liu, Y., Ren, X., & Cui, W. (2021). Design and implementation of a portable, low-power pulse oximeter based on STM32. Journal of Yan’an University (Natural Science Edition), 40(03), 79-82. https://doi.org/10.13876/J.cnki.ydnse.2021.03.079